Challenge Diabetes
Thoughts about current approaches to managing diabetes

Diabetes Intervention Technology™
Challenge Diabetes » Blog Archive » Worldwide Abbott Glucose Meter Correction Notice (FDA)

Worldwide Abbott Glucose Meter Correction Notice (FDA)

September 11th, 2007 by Kevin

Recall — Firm Press Release
Urgent: Abbott Notifies Users of Precision Xtra™, Optium™, ReliOn® Ultima, Rite Aid® and Kroger® Blood Glucose Meters to Check Display Screens

 

 

Not sure why it took me this long to find out about it?

 

 

From the FDA (also more info available at Abbott Diabetes):

 

 

FOR IMMEDIATE RELEASE — Alameda, CA — August 31, 2007 — Abbott is initiating a worldwide medical device correction for users of its Precision Xtra™, Optium™, ReliOn® Ultima, Rite Aid® and Kroger® blood glucose meters manufactured after January 31, 2007. If the meter is dropped on a hard surface, part or all of the display screen may not work properly or may appear blank, which could result in an inability to view blood glucose test results. The inability to generate blood glucose results could result in significant risk for hypoglycemia or hyperglycemia. Users of these meters who note that the display screen is not working properly should immediately stop using their meter as referenced in the User Guide and call Abbott Diabetes Care customer care for assistance at 1-877-844-4404. Through internal testing, Abbott Diabetes Care has found that when recently produced meters are dropped onto a hard surface, part of the display can be jarred or disconnected, thereby making it difficult to read the lot number or date information, or causing the screen to appear blank. No injuries have been reported to date.

Patients should keep their glucose meters in the wallet provided to offer additional protection for the meter. If the meter is dropped on a hard surface, patients should immediately perform a meter display check. Instructions on how to do this are detailed in the meter’s Users Guide. If no problems are encountered during the automatic display check, the meter is ready for use. Customers may call Abbott Diabetes Care customer care at 1-877-844-4404 to determine the date their meter was manufactured. Customers may also call customer care if they have questions or need a replacement meter.

Precision Xtra™, Optium, ReliOn® Ultima, Rite Aid® and Kroger® blood glucose meters have been distributed via retail and mail order pharmacies, physician offices and distributors. Abbott is notifying physicians, pharmacists, distributors and registered users by letter.

Blood glucose test strips used with these meters are not affected by this notification.

More information about this notification is available on the company’s website, www.abbottdiabetescare.com.

####

Contact:
Tama Antonia Donaldson
(510) 749-5449

 

FDA posts press releases and other notices of recalls and market withdrawals from the firms involved as a service to consumers, the media, and other interested parties. FDA does not endorse either the product or the company. This listserv covers mainly Class I (life-threatening) recalls. A complete listing of recalls can be found in the FDA Enforcement Report at: http://www.fda.gov/opacom/Enforce.html

 

FDA’s Recalls, Market Withdrawals and Safety Alerts Page: http://www.fda.gov/opacom/7alerts.html

To subscribe to or unsubscribe from this list, go to

https://list.nih.gov/archives/fda-recalls-l.html